Dry-well calibrators are portable instruments for the simple on-site calibration of temperature measuring instruments. A temperature dry-well calibrator consists of an electrically heated or cooled metal block, an accurate temperature control and an internal reference with indicator for determining the block temperature. In the metal block is really a bore for accepting the insert. This exchangeable insert is supplied with a number of bores with individual diameters and serves for the adaption of the test items.
With on-site calibration, the thermometer to be tested can be taken directly out of its thermowell in the plant, without having to interrupt the electrical connections to the evaluation system, and inserted into the appropriate bore of the insert. Here it will generally be ensured that the bore diameter, considering the thermal expansion of the calibration item, is selected to be as tight as you possibly can, since any air gap between your test item and the insert impairs heat transfer. Therefore, the bore diameter should be no more than 0.5 mm greater than the external diameter of the test item.
Structure of a dry-well calibrator
The required temperature with which the test item should be calibrated is regulated with the calibrator. Frequently Gargantuan are calibrated at the respective process temperature. Once a thermal equilibrium has been set between your thermometer being tested and the calibrator, the measuring deviation of the test item can be recorded. This corresponds to the temperature difference between its temperature and the reference temperature displayed on the calibrator. The total measurement uncertainty comprises of the accuracy of the calibrator and the short-term stability of the test item. Depending on the temperature range and the properties of the calibration item, calibration is carried out with a measurement uncertainty of 0.1 K to 3 K.
What else ought to be taken into account?
The minimum immersion depth for dry well calibrators is 70 mm. This is driven by the axial gradient, since warm air rises from underneath up. Should the minimum immersion depth not be reached, an external reference thermometer must be used for calibration. That is then placed on the same level as the object to be calibrated and the results are compared with one another.
Recommended guideline for the immersion depth of a test item
Diameter of the test item x 15 or for a known sensitive length: sensitive length + diameter of the test item x 10